YG6800 X-ray Fluorescence Multi-Element Analyzer is a new generation of microcomputerized benchtop analytical instrument. Breaking through the limitations of conventional domestic multi-element analyzers that only measure Al₂O₃, SiO₂, CaO, and Fe₂O₃, it can additionally analyze SO₃ and blended material content. This versatile five-element analyzer integrates the functionalities of XRF calcium-iron analyzers, silicon-aluminum analyzers, and sulfur analyzers, while automatically calculating the three key ratios: KH, SM, and IM. Comparative tests with imported large-scale X-ray fluorescence spectrometers show results within acceptable error margins. Purchasing one YG6800 can save nearly one million RMB, with low operational and maintenance costs, plus easy servicing.
Featuring a large LCD screen with Chinese menu prompts, the analyzer is user-friendly. Its integrated design ensures high reliability and easy maintenance. Equipped with a high-capacity electronic logbook, it stores over 2,000 analysis records, eliminating printing consumables. Samples remain intact for repeated testing. No chemical reagents are required, and the instrument produces no waste emissions, contains no radioactive sources, and has low power consumption—fully compliant with environmental protection, energy-saving, and radiation safety standards.
Applications:
- Measures CaO%, Fe₂O₃%, SiO₂%, Al₂O₃%, and SO₃% in raw materials, clinker, cement, limestone, clay, and gypsum.
- Determines SO₃%, CaO%, and blended material content in cement.
- Connects directly to raw material ratio batching control systems in the cement industry via RS232 serial communication.
Technical Specifications:
- Analysis Range: CaO, Fe₂O₃, SiO₂, Al₂O₃, SO₃: 0.01%–100%.
- Analysis Width: ≤15% for each component (CaO, Fe₂O₃, SiO₂, Al₂O₃, SO₃), determined via calibration curve.
- Analysis Precision (Standard Deviation):
- S CaO ≤ 0.10%
- S Fe₂O₃ ≤ 0.05%
- S SiO₂ ≤ 0.07%
- S Al₂O₃ ≤ 0.07%
- S SO₃ ≤ 0.05%
- Allowable Error:
- ΔCaO% ≤ 0.25%
- ΔFe₂O₃% ≤ 0.10%
- ΔSiO₂% ≤ 0.20%
- ΔAl₂O₃% ≤ 0.20%
- ΔSO₃% ≤ 0.15%
- Analysis Time: n × 180 seconds (n = 1, 2, 3; typically 2).
- Stability (Absolute Drift):
- ΔCaO% ≤ 0.15%
- ΔFe₂O₃% ≤ 0.10%
- ΔSiO₂% ≤ 0.10%
- ΔAl₂O₃% ≤ 0.10%
- ΔSO₃% ≤ 0.10%
- Operating Conditions:
- Power Supply: AC 200V–240V, 50Hz
- Power Consumption: ≤30W
- Ambient Temperature: 5–40°C
- Relative Humidity: ≤85% (at 30°C)
- Dimensions & Weight: 468mm × 368mm × 136mm, 13.8kg.