Sulfur-Calcium-Iron Analyzer Cement Plant X-Ray Fluorescence Sulfur-Calcium-Iron Analyzer This cement element testing instrument represents an upgraded successor to the newly launched calcium-iron analyzer. A single unit replaces both calcium-iron analyzers and sulfur testers, suitable for use both before grinding and after kilning in cement plants.
The KL3300 X-Ray Fluorescence Sulfur-Calcium-Iron Analyzer is a new microprocessor-based desktop instrument employing purely physical analysis methods. Designed for cement plants, it rapidly measures the percentage content of SO₃, CaO, and Fe₂O₃ in raw materials, clinker, cement, gypsum, etc., within 30 seconds. Its high-speed analysis enables real-time monitoring of changes in these three components during production, facilitating timely adjustments to raw material ratios. This lays a solid foundation for producing qualified clinker and cement. Specifically, determining the CaO percentage in finished cement enables cement manufacturers to produce qualified cement of a specific grade without overusing or underusing clinker, generating significant economic benefits for cement enterprises. Measuring the SO₃ and CaO percentages in finished cement allows manufacturers to predict cement strength and the setting time of concrete mixed with that cement.
Featuring a large LCD screen with fully Chinese-language menu prompts and intelligent error alerts, the instrument offers exceptional user convenience.
Analysis Accuracy:
Standard Deviation
S₂O₃ ≤ 0.04%
S₂CaO ≤ 0.10%
S₂Fe₂O₃ ≤ 0.05%
Rapid analysis delivers percentage content results for SO₃, CaO, and Fe₂O₃ within 30 seconds.
Requires no chemical reagents, produces no waste emissions, contains no radioactive sources, and features low power consumption—meeting environmental, energy-saving, and safety requirements.

X-Ray Fluorescence Sulfur-Calcium-Iron Analyzer Technical Specifications:
- Analysis Range: SO₃, CaO, Fe₂O₃: 0.01%~*All ranges are adjustable.
- Analysis Range Width: SO₃ (CaO, Fe₂O₃) %max~SO₃ (CaO, Fe₂O₃) %min ≤ 15% CaO min, determined by establishing calibration curves.
- Permissible Error: SO₃: ±0.11%, CaO: ±0.18%, Fe₂O₃: ±0.10%.
- System Analysis Time: 30×n seconds (n is a natural number), recommended value: 60 seconds.
- Analysis Accuracy: S SO₃ ≤ 0.04%, S CaO ≤ 0.10%, S Fe₂O₃ ≤ 0.05%.
- Temperature Stability: Within 5~40℃ range, drift: |ΔSO₃| ≤ 0.10 |ΔCaO| ≤ 0.10 |ΔFe₂O₃| ≤ 0.10%.
- Operating Conditions: Ambient temperature: 5~40℃, relative humidity: ≤85% (at 30℃), Power supply: 220V ±20V, 50Hz.
- Total power consumption: ≤50W.
- Dimensions and weight: 468×368×136mm, 13.8kg